5522A
Operators Manual
1-6
Remote Operation (RS-232)
There are two rear-panel serial data RS-232 ports: SERIAL 1 FROM HOST, and
SERIAL 2 TO UUT (see Figure 1-2). Each port is dedicated to serial data
communications for operating and controlling the 5522A during calibration procedures.
For complete information on remote operations, see Chapter 5.
The SERIAL 1 FROM HOST serial data port connects a host terminal or personal
computer to the Calibrator. You have several choices for sending commands to the
Calibrator: you can enter commands from a terminal (or a PC running a terminal
program), you can write your own programs using BASIC, or you can run optional
Windows-based software such as 5500/CAL or MET/CAL. The 5500/CAL Software
includes more than 200 example procedures covering a wide range of test tools the
5522A can calibrate. (See Chapter 6 for a discussion of the RS-232 commands.)
The SERIAL 2 TO UUT serial data port connects a UUT to a PC or terminal via the
5522A (see Figure 1-2). This “pass-through” configuration eliminates the requirement for
two COM ports at the PC or terminal. A set of four commands control the operation of
the SERIAL 2 TO UUT serial port. See Chapter 6 for a discussion of the UUT_*
commands. The SERIAL 2 TO UUT port is also used to connect to the Fluke 700 Series
Pressure Modules.
0
•
1 2 3
4 5 6
7
8 9
ENTER
M
k
m
V Hz
/
+
F
OPRSTBY
EARTH SCOPE
EXGRD
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
n
p
W
dBm sec
¡C
POWER
A
MULT
x
DIV
MODES
MORE
5522A
CALIBRATOR
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
SCOPE
OUT
V,
, ,
RTD
A, -SENSE,
AUX V
20V PK MAX
20V PK MAX
FIELD
EDIT
5522A
CALIBRATOR
0 •
123
456
7
8 9
ENTER
M
k
m
V
Hz
/
+
F
OPRSTBY EARTH
SCOPE
EXGRD
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
n
p
W
dBm
sec
¡C
POWER
A
MULT
x
DIV
MODES
MORE
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
20V PK MAX
20V PK MAX
FIELD
EDIT
5522A
CALIBRATOR
Unit Under Test
5522A
5522A
SERIAL 1 FROM HOST port
COM port
RS-232 Remote Operation using the
SERIAL 1 FROM HOST port
SERIAL 1 FROM HOST port
SERIAL 2
TO UUT port
RS-232 Remote Operation using the
SERIAL 1 FROM HOST and
SERIAL 2 TO UUT ports
PC or Terminal
COM port
PC or Terminal
gjh002.eps
Figure 1-2. RS-232 Remote Connection
Remote Operation (IEEE-488)
The rear panel IEEE-488 port is a fully programmable parallel interface bus meeting
standard IEEE-488.1 and supplemental standard IEEE-488.2. Under the remote control of
an instrument controller, the Calibrator operates exclusively as a “talker/listener.” You