SC1100 Oscilloscope Calibration Option
General Specifications10
10-5
Edge Specifications
Edge Characteristics into 50 Ω Load
1-Year Absolute Uncertainty,
tcal ± 5 °C
Rise Time
300 ps
(+0 ps / -100 ps)
Amplitude Range (p-p) 5.0 mV to 2.5 V
±(2 % of output + 200 μV)
Resolution 4 digits n/a
Adjustment Range
±10 % around each sequence value
(indicated below)
n/a
Sequence Values
5 mV, 10 mV, 25 mV, 50 mV, 60 mV, 80
mV, 100 mV, 200 mV, 250 mV, 300 mV,
500 mV, 600 mV, 1 V, 2.5 V
n/a
Frequency Range 1 kHz to 10 MHz
[1]
±(2.5 ppm of setting)
Typical Jitter, edge to trigger <5 ps (p-p) n/a
within 2 ns from 50 % of rising edge <(3 % of output + 2 mV)
2 to 5 ns <(2 % of output + 2 mV)
5 to 15 ns <(1 % of output + 2 mV)
Leading Edge Aberrations
[2]
after 15 ns <(0.5 % of output + 2 mV)
Typical Duty Cycle 45 % to 55 % n/a
Tunnel Diode Pulse Drive Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 V to 100 V p-p.
[1] Above 2 MHz, the rise time specification is <350 ps.
[2] All edge aberration measurements are made with a Tektronix 11801 mainframe with an SD26 input module.
Leveled Sine Wave Specifications
Leveled Sine Wave Frequency Range
Characteristics
into 50 Ω
50 kHz (reference) 50 kHz to 100 MHz 100 to 300 MHz 300 to 600 MHz 600 to 1100 MHz
Amplitude Characteristics (for measuring oscilloscope bandwidth)
Range (p-p) 5 mV to 5.5 V 5 mV to 3.5 V
Resolution
<100 mV: 3 digits
100 mV: 4 digits
Adjustment
Range
continuously adjustable
1-Year Absolute
Uncertainty,
tcal ± 5 °C
±(2 % of output
+ 300 μV)
±(3.5 % of output
+ 300 μV)
±(4 % of output
+ 300 μV)
±(6 % of
output
+ 300 μV)
±(7 % of output
+ 300 μV)
Flatness (relative
to 50 kHz)
not applicable
±(1.5 % of output
+ 100 μV)
±(2 % of output
+ 100 μV)
±(4 % of
output
+ 100 μV)
±(5 % of output
± 100 μV)
Short-Term
Amplitude
Stability
1 %
[1]
Frequency Characteristics
Resolution 10 kHz 100 kHz
1-Year Absolute
Uncertainty,
tcal ± 5 °C
±2.5 ppm
[2]
Distortion Characteristics
2nd Harmonic
-33 dBc
3rd and Higher
Harmonics
-38 dBc
[1] Within one hour after reference amplitude setting, provided temperature varies no more than ±5 °C.
[2] With REF CLK set to ext, the frequency uncertainty of the Leveled Sine Wave is the uncertainty of the external 10 MHz clock
±0.3 Hz/gate time.