5522A
Operators Manual
10-4
Calibration (tcal).................................................. 15 °C to 35 °C
Storage ............................................................... -20 °C to +70 °C
Electromagnetic Compatibility ............................ Designed to operate in Standard Laboratory environments where the
Electromagnetic environment is highly controlled. If used in areas with
Electromagnetic fields >1 V/m, there could be errors in output values.
All testing for this specification used new cables and connectors.
Temperature Coefficient....................................... Temperature Coefficient for temperatures outside tcal +5 °C is
10 % per °C of 1-year specification.
Relative Humidity
Operating ............................................................ <80 % to 30 °C, <70 % to 40 °C,<40 % to 50 °C
Storage ............................................................... <95 %, noncondensing
Altitude
Operating ............................................................ 3,050 m (10,000 ft) maximum
Nonoperating ...................................................... 12,200 m (40,000 ft) maximum
Safety ..................................................................... Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation............................................ 20 V
EMC ........................................................................ Complies with EN 61326-1/1997, Class A
Volt Specifications
DC Signal
Square Wave Signal
[1]
Volt Function
50 Ω Load 1 MΩ Load 50 Ω Load 1 MΩ Load
Amplitude Characteristics
Range
0 to ±6.6 V 0 to ±130 V
±1 mV to
±6.6 V p-p
±1 mV to
±130 V p-p
Resolution
Range
1 to 24.999 mV
25 to 109.99 mV
110 mV to 2.1999 V
2.2 to 10.999 V
11 to 130 V
Resolution
1 μV
10 μV
100 μV
1 mV
10 mV
Adjustment Range Continuously adjustable
1-Year Absolute Uncertainty, tcal ±
5 °C
±(0.25 % of
output + 40 μV)
±(0.05 % of
output + 40 μV)
±(0.25 % of
output + 40 μV)
±(0.1% of output +
40 μV)
[2]
Sequence
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range
10 Hz to 10 kHz
1-Year Absolute Uncertainty, tcal ±
5 °C
±(2.5 ppm of setting)
Typical Aberration
within 4 μs from 50 % of
leading/trailing edge
<(0.5 % of output + 100 μV)
[1] Selectable positive or negative, zero referenced square wave.
[2] For square wave frequencies above 1 kHz, ±(0.25 % of output + 40 μV).