SC1100 Oscilloscope Calibration Option
Verification Tables10
10-41
Table 10-8. SC1100 Option Wave Generator Amplitude Verification (50 output impedance) (cont.)
Wave Shape
Nominal
Value (V p-p)
Frequency
(Hz)
Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
sine 0.0449 1000 0.001447
sine 0.109 1000 0.00337
sine 0.449 1000 0.01357
sine 1.09 1000 0.0328
sine 2.5 1000 0.0751
triangle 0.0018 1000 0.000154
triangle 0.0109 1000 0.000427
triangle 0.0449 1000 0.001447
triangle 0.109 1000 0.00337
triangle 0.449 1000 0.01357
triangle 1.09 1000 0.0328
triangle 2.5 1000 0.0751
Edge Verification: Amplitude
Table 10-9. SC1100 Option Edge Verification: Amplitude
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.005 1 kHz 0.0003
0.005 10 kHz 0.0003
0.005 100 kHz 0.0003
0.01 100 kHz 0.0004
0.025 100 kHz 0.0007
0.05 100 kHz 0.0012
0.1 100 kHz 0.0022
0.25 100 kHz 0.0052
0.5 100 kHz 0.0102
1 100 kHz 0.0202
2.5 100 kHz 0.0502
2.5 10 kHz 0.0502
2.5 1 kHz 0.0502