PQ Option
5522A PQ Option Specifications11
11-5
AC Voltage Auxiliary Specifications (Dual Output Mode Only)
Range,
Composite
Waveform
Harmonic
Frequency
Harmonic
Amplitude
Range ( % of
Fundamental)
Harmonic Amplitude
Uncertainty
(% of Fundamental + V)
Harmonic Phase
Uncertainty
(Relative to
Fundamental)
[1]
Absolute RMS
Uncertainty of
Composite Waveform
(% RMS + V)
15 to 45 Hz 0.1 to 100 %
0.1 % + 100 μV 0.5 °
45 Hz to 1 kHz 0.1 to 100 %
0.1 % + 100 μV 1 °
1 to 2 kHz 0.1 to 50 %
0.1 % + 100 μV 3 °
10 to
329.99 mV
2 to 5 kHz 0.1 to 30 %
0.1 % + 500 μV 6 °
0.2 % + 100 μV
15 to 45 Hz 0.1 to 100 % 0.1 % + 1 mV
0.5 °
45 Hz to 1 kHz 0.1 to 100 % 0.1 % + 1 mV
0.75 °
1 to 2 kHz 0.1 to 50 % 0.1 % + 1 mV
2 °
.33 to
3.2999 V
2 to 5 kHz 0.1 to 30 % 0.1 % + 2 mV
3 °
0.2 % + 1 mV
15 to 45 Hz 0.1 to 100 % 0.2 % + 3 mV
0.5 °
45 Hz to 1 kHz 0.1 to 100 % 0.2 % + 3 mV
0.75 °
1 to 2 kHz 0.1 to 50 % 0.2 % + 3 mV
2 °
3.3 to 5 V
2 to 5 kHz 0.1 to 30 % 0.3 % + 3 mV
3 °
0.2 % + 2 mV
[1] For harmonics that are < 1 % of the Fundamental, phase uncertainty is typical.
AC Current Specifications, LCOMP OFF
Range,
Composite
Waveform
Harmonic
Frequency
Harmonic
Amplitude
Range
(% of
Fundamental)
[1]
Harmonic Amplitude
Uncertainty
(% of
Fundamental + A)
Harmonic Phase
Uncertainty
(Relative to
Fundamental)
[2]
Absolute RMS
Uncertainty of
Composite
Waveform
(% RMS + A)
15 to 45 Hz 0.1 to 100 %
0.1 % + 0.1 μA 0.5 °
45 to 900 Hz 0.1 to 100 %
0.1 % + 0.1 μA 2 °
900 Hz to 2 kHz 0.1 to 50 %
0.1 % + 0.1 μA 3 °
29 to
329.9 μA
2 to 5 kHz 0.1 to 30 %
0.1 % + 0.13 μA 6 °
0.2 % + 0.1 μA
15 to 45 Hz 0.1 to 100 %
0.1 % + 1 μA 0.5 °
45 to 900 Hz 0.1 to 100 %
0.1 % + 1 μA 0.6 °
900 Hz to 2 kHz 0.1 to 50 %
0.1 % + 1 μA 0.75 °
0.33 to
3.299 mA
2 to 5 kHz 0.1 to 30 %
0.1 % + 1.3 μA 2 °
0.2 % + 1 μA
15 to 45 Hz 0.1 to 100 %
0.1 % + 10 μA 0.5 °
45 to 900 Hz 0.1 to 50 %
0.1 % + 10 μA 0.6 °
900 Hz to 2 kHz 0.1 to 30 %
0.1 % + 10 μA 0.75 °
3.3 to 32.99 mA
2 to 5 kHz 0.1 to 100 %
0.1 % + 13 μA 2 °
0.2 % + 10 μA
15 to 45 Hz 0.1 to 100 %
0.1 % + 100 μA 0.5 °
45 to 900 Hz 0.1 to 100 %
0.1 % + 100 μA 0.75 °
900 Hz to 2 kHz 0.1 to 50 %
0.1 % + 100 μA 1.5 °
33 to 329.9 mA
2 to 5 kHz 0.1 to 30 %
0.1 % + 130 μA 3 °
0.2 % + 100 μA
[1] All frequencies can have harmonics up to 100 % of the fundamental; uncertainties are not specified unless otherwise indicated.
[2] For harmonics that are < 1 % of the Fundamental, phase uncertainty is typical.