EasyManua.ls Logo

Fluke 5522A - Verification Tables; DC Voltage Verification

Fluke 5522A
426 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SC1100 Oscilloscope Calibration Option
Verification Tables10
10-35
Verification Tables
The verification test points are provided here as a guide when verification to one-year
specifications is desired.
DC Voltage Verification
Table 10-2. SC1100 Option DC Voltage Verification
(1 MΩ output impedance unless noted)
Nominal
Value (V dc)
Measured Value
(V dc)
Deviation
(V dc)
1-Year Spec. (V dc)
0 0.00004
0.00125 0.000040625
-0.00125 0.000040625
0.00249 0.000041245
-0.00249 0.000041245
0.0025 0.00004125
-0.0025 0.00004125
0.00625 0.000043125
-0.00625 0.000043125
0.0099 0.00004495
-0.0099 0.00004495
0.01 0.000045
-0.01 0.000045
0.0175 0.00004875
-0.0175 0.00004875
0.0249 0.00005245
-0.0249 0.00005245
0.025 0.0000525
-0.025 0.0000525
0.0675 0.00007375
-0.0675 0.00007375
0.1099 0.00009495
-0.1099 0.00009495
0.11 0.000095
-0.11 0.000095
0.305 0.0001925
-0.305 0.0001925