Loading...
5522A
Operators Manual
10-36
Table 10-2. SC1100 Option DC Voltage Verification (cont.)
Nominal
Value (V dc)
Measured Value
(V dc)
Deviation
(V dc)
1-Year Spec. (V dc)
0.499 0.0002895
-0.499 0.0002895
0.5 0.00029
-0.5 0.00029
1.35 0.000715
-1.35 0.000715
2.19 0.001135
-2.19 0.001135
2.2 0.00114
-2.2 0.00114
6.6 0.00334
-6.6 0.00334
10.99 0.005535
-10.99 0.005535
11 0.00554
-11 0.00554
70.5 0.03529
-70.5 0.03529
130 0.06504
-130 0.06504
Table 10-3. SC1100 Option DC Voltage Verification at 50 Ω
Calibrator
Mainframe
Output
Measured Value
(V DC)
Reading x correction Tolerance (V DC)
0 mV 0.00004 V
2.49 mV 4.623E-05 V
-2.49 mV 4.623E-05 V
9.90 mV 6.475E-05 V
-9.90 mV 6.475E-05 V
24.9 mV 0.0001023 V
-24.9 mV 0.0001023 V
109.9 mV 0.0003148 V

Table of Contents