SC1100 Oscilloscope Calibration Option
Verification Tables10
10-37
Table 10-3. SC1100 Option DC Voltage Verification at 50 Ω (cont.)
Calibrator
Mainframe
Output
Measured Value
(V DC)
Reading x correction Tolerance (V DC)
-109.9 mV 0.0003148 V
499 mV 0.0012875 V
-499 mV 0.0012875 V
2.19 V 0.005515 V
-2.19 V 0.005515 V
6.599 V 0.0165375 V
-6.599 V 0.0165375 V
AC Voltage Verification
Table 10-4. SC1100 Option AC Voltage Verification
(1 MΩ output impedance unless noted)
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-year Spec.
(V p-p)
0.001 1000 0.000041
-0.001 1000 0.000041
0.01 1000 0.00005
-0.01 1000 0.00005
0.025 1000 0.000065
-0.025 1000 0.000065
0.11 1000 0.00015
-0.11 1000 0.00015
0.5 1000 0.00054
-0.5 1000 0.00054
2.2 1000 0.00224
-2.2 1000 0.00224
11 1000 0.01104
-11 1000 0.01104
130 1000 0.13004
-130 1000 0.13004
200 mV 100 0.00024
200 mV 1000 0.00024
200 mV 5000 0.00054