5522A
Operators Manual
10-32
RANGE
(IEEE-488, RS-232, Sequential)
Programs the instrument range in PULSE, MEAS Z, OVERLD modes.
Parameters:
Pulse
TP0DB TP8DB TP20DB TP28DB TP40DB TP48DB
Range 2.5 V 1.0 V 250 mV 100 mV 25 mV 10 mV
Impedance
Measure
TZ50OHM TZ1MOHM TZCAP
Range res 50 Ω res 1MΩ cap
Impedance
Measure
TZ50OHM TZ1MOHM TZCAP
Range res 50 Ω res 1MΩ cap
Example: RANGE TP20DB
Edge Function Commands
TDPULSE
(IEEE-488, RS-232, Sequential)
Turns tunnel diode pulse drive on/off in EDGE mode.
Parameters: ON (or non-zero) or OFF (or zero)
Example: TDPULSE ON
Returns the tunnel diode pulse drive setting in EDGE mode.
Parameters: None
Response: 1 if ON, 0 if OFF.
Marker Function Commands
TMWAVE
(IEEE-488, RS-232, Sequential)
Selects the waveform for MARKER mode.
Parameters: SINE Sine wave (2 ns to 15 ns)
SPIKE Triangular/sawtooth pulse (15 ns to 5s)
SQUARE Square wave (50 % duty cycle) (4 ns to 5s)
SQ20PCT Square wave (20 % duty cycle) (85 ns to 5s)
Example: TMWAVE SPIKE