SC1100 Oscilloscope Calibration Option
Verification Tables10
10-43
Tunnel Diode Pulser Verification
Table 10-13. SC1100 Option Tunnel Diode Pulser Verification
Nominal Value
(V p-p)
Frequency (Hz) Measured Value
(V p-p)
Deviation (V p-p) 1-Year Spec.
(V p-p)
11 100 0.2202
11 10000 0.2202
55 100 1.1002
55 10000 1.1002
100 100 2.0002
100 10000 2.0002
Leveled Sinewave Verification: Amplitude
Table 10-14. SC1100 Option Leveled Sinewave Verification: Amplitude
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.005 50 kHz 0.0004
0.0075 50 kHz 0.00045
0.0099 50 kHz 0.000498
0.01 50 kHz 0.0005
0.025 50 kHz 0.0008
0.039 50 kHz 0.00108
0.04 50 kHz 0.0011
0.07 50 kHz 0.0017
0.099 50 kHz 0.00228
0.1 50 kHz 0.0023
0.25 50 kHz 0.0053
0.399 50 kHz 0.00828
0.4 50 kHz 0.0083
0.8 50 kHz 0.0163
1.2 50 kHz 0.0243
1.3 50 kHz 0.0263
3.4 50 kHz 0.0683
5.5 50 kHz 0.1103