SC600 Oscilloscope Calibration Option
Verification Tables 9
9-33
Verification Tables
The verification test points are provided here as a guide when verification to one-year
specifications is desired.
DC Voltage Verification
Table 9-2. SC600 Option DC Voltage Verification
(1 MΩ output impedance unless noted)
Nominal Value (V dc) Measured Value (V dc) Deviation (V dc) 1-Year Spec. (V dc)
0 0.00004
0.00125 0.000040625
-0.00125 0.000040625
0.00249 0.000041245
-0.00249 0.000041245
0.0025 0.00004125
-0.0025 0.00004125
0.00625 0.000043125
-0.00625 0.000043125
0.0099 0.00004495
-0.0099 0.00004495
0.01 0.000045
-0.01 0.000045
0.0175 0.00004875
-0.0175 0.00004875
0.0249 0.00005245
-0.0249 0.00005245
0.025 0.0000525
-0.025 0.0000525
0.0675 0.00007375
-0.0675 0.00007375
0.1099 0.00009495
-0.1099 0.00009495
0.11 0.000095
-0.11 0.000095
0.305 0.0001925
-0.305 0.0001925
0.499 0.0002895