5522A
Operators Manual
10-40
Table 10-7. SC1100 Option Wave Generator Amplitude Verification (1 M output impedance) (cont.)
Wave Shape
Nominal
Value (V p-p)
Frequency
(Hz)
Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
triangle 0.219 1000 0.00667
triangle 0.899 1000 0.02707
triangle 6.59 1000 0.1978
triangle 55 1000 1.6501
Wave Generator Amplitude Verification: 50
Ω
Output Impedance
Table 10-8. SC1100 Option Wave Generator Amplitude Verification (50 output impedance)
Wave Shape
Nominal
Value (V p-p)
Frequency
(Hz)
Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
square 0.0018 1000 0.000154
square 0.0064 1000 0.000292
square 0.0109 1000 0.000427
square 0.011 1000 0.00043
square 0.028 1000 0.00094
square 0.0449 1000 0.001447
square 0.045 1000 0.00145
square 0.078 1000 0.00244
square 0.109 1000 0.00337
square 0.11 1000 0.0034
square 0.28 1000 0.0085
square 0.449 1000 0.01357
square 0.45 1000 0.0136
square 0.78 1000 0.0235
square 1.09 1000 0.0328
square 1.1 1000 0.0331
square 1.8 1000 0.0541
square 2.5 10 0.0751
square 2.5 100 0.0751
square 2.5 1000 0.0751
square 2.5 10000 0.0751
sine 0.0018 1000 0.000154
sine 0.0109 1000 0.000427