SC600 Oscilloscope Calibration Option
How to Test the Trigger SC600 Option 9
9-21
Peaks are aligned
with center axis
gl011i.eps
4. Repeat this procedure for all time marker values recommended for your oscilloscope.
Repeat for digital and analog mode as required. Some oscilloscopes may need the
magnification changed while calibrating in analog mode.
5. Remove the input signal by pressing Y.
How to Test the Trigger SC600 Option
The oscilloscope’s ability to trigger on different waveforms can be tested using the wave
generator. When the wave generator is used, a square, sine, or triangle wave is
transmitted and the wave’s output impedance, offset, and voltage can be varied in order
to test the triggering capability at different levels.
NOTE
The wave generator should not be used for checking the accuracy of your
oscilloscope.
The wave generator is available through the Wavegen menu, shown below. To access this
menu, press the softkey under MODE until “wavegen” appears.
gjh061.eps
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P to return directly to the OTHER modes menu.
Each option in the Wavegen menu is described below.