SC600 Oscilloscope Calibration Option
Oscilloscope Connections 9
9-7
Overload Measurement Specifications
Source Voltage
Typical ‘On’ Current
Indication
Typical ‘Off’ Current Indication
Maximum Time Limit DC or AC
(1 kHz)
5 to 9 V 100 to 180 mA 10 mA Setable 1 s to 60 s
Oscilloscope Connections
Using the cable supplied with the SC600 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 9-1).
To use the external trigger, connect the TRIG OUT output on the Calibrator to the
external trigger connection on your oscilloscope. To use the external trigger and view its
signal with the calibration signal, connect the TRIG OUT output to another channel. See
your oscilloscope manual for details on connecting and viewing an external trigger.
5522A
CALIBRATOR
gjh036.eps
Figure 9-1. Oscilloscope Connection: Channel and External Trigger
How to Start the SC600 Option
Press a (LED lit) to select the SC600 Option. The SCOPE menu, shown below,
appears in the Control Display. You can press any of the first four softkeys to go directly
to the VOLT, EDGE, LEVSINE, and MARKER calibration menus. Press the last softkey
to go to the OTHER menu (also shown below), allowing access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Press P to return to the SCOPE menu from the OTHER menu. This chapter
describes each of these menus in detail.