SC600 Oscilloscope Calibration Option
Verification Tables 9
9-49
Edge Verification: Duty Cycle
Table 9-13. SC600 Option Edge Verification: Duty Cycle
Nominal Value
(V p-p)
Frequency
Measured Value
(%)
Deviation
(from 50%)
1-Year Spec. (%)
2.5 1 MHz 5
Edge Verification: Rise Time
Table 9-14. SC600 Option Edge Verification: Rise Time
Nominal Value
(V p-p)
Frequency
Measured Value
(s)
Deviation (ns)
1-Year Spec.
(ns)
0.25 1 kHz 0.3 ns
0.25 100 kHz 0.3 ns
0.25 10 MHz 0.3 ns
0.5 1 kHz 0.3 ns
0.5 100 kHz 0.3 ns
0.5 10 MHz 0.3 ns
1 1 kHz 0.3 ns
1 100 kHz 0.3 ns
1 10 MHz 0.3 ns
2.5 1 kHz 0.3 ns
2.5 100 kHz 0.3 ns
2.5 10 MHz 0.3 ns
Tunnel Diode Pulser Verification
Table 9-15. SC600 Option Tunnel Diode Pulser Verification
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec. (V p-
p)
11 100 0.2202
11 10000 0.2202
55 100 1.1002
55 10000 1.1002
100 100 2.0002
100 10000 2.0002