4-54
5522A #2
Aux
Aux
LO
20A
LO
20A
10 MHz
IN
10 MHz
OUT
Load/Meter
5522A #1
gjh023.eps
Figure 4-16. Two Calibrators Sourcing Current in Parallel
Three-Phase Power Calibration
You can configure three Calibrators to calibrate a three-phase power meter. This example
uses the assumption that you want to apply a perfectly balanced calibration output with a
unity power factor. By changing the phase relationships, you can apply other test
stimulus. The figure shows the phase relationship of each Calibrator. By changing the
phase relationships, you can apply other test stimulus.