5522A
Operators Manual
1-24
AC Voltage, DC Offset
Range
[1]
(Normal Channel) Offset Range
[2]
Max Peak
Signal
1-Year Absolute Uncertainty,
tcal ±5 °C
[3]
±(% dc output + floor)
Sine Waves (rms)
3.3 to 32.999 mV 0 to 50 mV 80 mV
0.1 + 33 μV
33 to 329.999 mV 0 to 500 mV 800 mV
0.1 + 330 μV
0.33 to 3.29999 V 0 to 5 V 8 V
0.1 + 3300 μV
3.3 to 32.9999 V 0 to 50 V 55 V 0.1 + 33 mV
Triangle Waves and Truncated Sine Waves (p-p)
9.3 to 92.999 mV 0 to 50 mV 80 mV
0.1 + 93 μV
93 to 929.999 mV 0 to 500 mV 800 mV
0.1 + 930 μV
0.93 to 9.29999 V 0 to 5 V 8 V
0.1 + 9300 μV
9.3 to 93.0000 V 0 to 50 V 55 V 0.1 + 93 mV
Square Waves (p-p)
6.6 to 65.999 mV 0 to 50 mV 80 mV
0.1 + 66 μV
66 to 659.999 mV 0 to 500 mV 800 mV
0.1 + 660 μV
0.66 to 6.59999 V 0 to 5 V 8 V
0.1 + 6600 μV
6.6 to 66.0000 V 0 to 50 V 55 V 0.1 + 66 mV
[1] Offsets are not allowed on ranges above the highest range shown above.
[2] The maximum offset value is determined by the difference between the peak value of the selected voltage output and the
allowable maximum peak signal. For example, a 10 V p-p square wave output has a peak value of 5 V, allowing a maximum
offset up to
± 50 V to not exceed the 55 V maximum peak signal. The maximum offset values shown above are for the
minimum outputs in each range.
[3] For frequencies 0.01 to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5 % of output,
±1 % of the offset range.
AC Voltage, Square Wave Characteristics
Risetime @
1 kHz
Typical
Settling Time @
1 kHz Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty
<1 μs
<10 μS to 1 % of
final value
<2 %
1 % to 99 % <3.3 V p-p.
0,01 Hz to 100 kHz
±(0.02 % of period + 100 ns), 50 % duty cycle
±(0.05 % of period + 100 ns), other duty cycles
from 10 % to 90 %
±(0.8 % of period +100 ns)
AC Voltage, Triangle Wave Characteristics
Linearity to 1 kHz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range
AC Current (Non-Sine Wave)
Triangle Wave &
Truncated Sine Wave
Range
p-p
Frequency
1-Year Absolute Uncertainty tcal ±5 °C
±(% of output + % of range)
Max Current
Resolution
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
0.047 to
0.92999 mA
[1]
1 to 10 kHz 10 + 2
Six digits
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
0.93 to
9.29999 mA
[1]
1 to 10 kHz 10 + 2
Six digits
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
9.3 to
92.9999 mA
[1]
1 to 10 kHz 10 + 2
Six digits