Introduction and Specifications
Additional Specifications 1
1-23
AC Voltage (Non-Sine Wave) (cont.)
Triangle Wave &
Truncated Sine
Range, p-p
[1]
Frequency
1-Year Absolute Uncertainty,
tcal ±5 °C,
±(% of output + % of range)
[2]
Max Voltage
Resolution
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
9.3 to 14.0000 V
1 to 10 kHz 5.0 + 0.5
Six digits on each range
[1] To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for truncated sine wave,
multiply the p-p value by 0.2165063.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Uncertainty for Truncated Sine outputs is typical over this frequency band.
Square Wave
Range
(p-p)
[1]
Frequency
1-Year Absolute Uncertainty,
tcal ±5 °C,
±(% of output + % of range)
[2]
Max Voltage
Resolution
Normal Channel (Single Output Mode)
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
2.9 to 65.999 mV
20 to 100 kHz 5.0 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
66 to 659.999 mV
20 to 100 kHz 5.0 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
0.66 to 6.59999 V
20 to 100 kHz 5.0 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
1 to 20 kHz 0.5 + 0.25
6.6 to 66.0000 V
20 to 100 kHz 5.0 + 0.5
Six digits on each range
Auxiliary Output (Dual Output Mode)
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
29 to 659.999 mV
1 to 10 kHz
[3]
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
0.66 to 6.59999 V
1 to 10 kHz
[3]
5.0 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5 Two digits on each range
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
6.6 to 14.0000 V
1 to 10 kHz
[3]
5.0 + 0.5
Six digits on each range
[1] To convert p-p to rms for square wave, multiply the p-p value by 0.5.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3] Limited to 1 kHz for Auxiliary outputs
6.6 V p-p.