PQ Option
Delta (() Amplitude, Single (Sags & Swells) Function (Current)11
11-27
RAMPTIME?
(IEEE-488, RS-232, Sequential)
Read the current Ramptime setting.
Example: RAMPTIME?
_______________________________________________________________________
RPTFREQ <MOD_FREQ>
(IEEE-488, RS-232, Coupled)
Set the modulation frequency in Flicker mode.
<MOD_FREQ> 0.001Hz to 30.0Hz
Example: RPTFREQ 10Hz
Example: RPTFREQ 5
_______________________________________________________________________
RPTFREQ?
(IEEE-488, RS-232, Sequential)
Return the modulation frequently in Flicker mode.
Example: RPTFREQ?
_______________________________________________________________________
TRIGGED?
(IEEE-488, RS-232, Sequential)
Returns 0 if no trig even, 1 if triggered
_______________________________________________________________________
Example Strings
The following sets up a Flicker output on the primary channel, with a square wave
modulation frequency of 0.65 Hz and a delta V/V of 0.91 %. The output voltage is 120 V,
60 Hz:
pq dampl; evtmode repeat; flickwav square; rptfreq .65;
deltamag pri, .0091; out 120v, 60hz; oper