11-26
PQ <MODE>
(IEEE-488, RS-232, Coupled)
Enter/Exit the Power Quality mode.
<MODE> is OFF --- for normal 5522 mode
CH --- for Composite Harmonic mode
DAMPL --- for Sag/Swell or Flicker mode
Example: PQ CH --- Enter the Composite Harmonics mode
_______________________________________________________________________
PQ?
(IEEE-488, RS-232, Sequential)
Returns the Power Quality calibration mode.
OFF --- currently operating in normal 5522 mode
CH --- currently in composite Harmonic mode
DAMPL --- currently in Sag/Swell or Flicker mode
Example: PQ?
_______________________________________________________________________
POSTTRIGDELAY <D_TIME>
(IEEE-488, RS-232, Coupled)
Set a post trigger delay in Single (Sag/Swell) mode.
<D_TIME> is 0.001 Sec to 60 Sec
Example: POSTTRIGDELAY 3
_______________________________________________________________________
POSTTRIGDELAY?
(IEEE-488, RS-232, Sequential)
Read the post trigger delay in Single (Sag/Swell) mode.
Example: POSTTRIGDELAY?
_______________________________________________________________________
RAMPTIME <R_TIME>
(IEEE-488, RS-232, Coupled)
Set the period over which the output amplitude is "ramped" up/down to the selected Delta
V / V value in Single (Sag/Swell) mode.
<R_TIME> is 0.001 Sec to 60.0 Sec
Example: RAMPTIME 5.3
_______________________________________________________________________