PQ Option
11
11-7
Flicker Simulation Mode
Voltage Range 1 mV to 1020 V
Current Range
29 μA to 20.5 A
Frequency of Fundamental 50 and 60 Hz
Amplitude Modulation Range
±100 %
Frequency of Modulation 0.1 to 40 Hz
Type of Modulation Square or Sine
Short Term (10 minute) uncertainty of amplitude modulation
±0.1 % of nominal output + 0.05% of range
Flicker Modulation Timing Uncertainty
±0.1 ms
Settings for Pst = 1
Voltage Changes ΔV/V %
[1]
Changes per minute: 120V, 60 Hz 230V, 50 Hz
1 chg/min 3.166 % 2.724 %
2 chg/min 2.568 % 2.211 %
7 chg/min 1.695 % 1.459 %
39 chg/min 1.044 % 0.906 %
110 chg/min 0.841 % 0.725 %
1620 chg/min 0.547 % 0.402 %
4000 chg/min N/A 2.40 %
4800 chg/min 3.920 % N/A
Trigger Event
2
nd
Push of OPER key, or Remote Command
[1] Values shown are nominal values per IEC 61000-4-15. The 5520A/PQ has a limited resolution of 0.02 % in the Flicker Simulation
Mode.
Sags & Swells Simulation Mode
Voltage Range
1 mV to 1020 V
Current Range 29 μA to 20.5 A
Frequency of Fundamental
45 to 65 Hz
Amplitude Modulation Range ±100 %
Ramp-Up Time
0.01 to 1 second
Duration of Sag or Swell
0.032 to 60 seconds
Trigger Event
2
nd
Push of OPER key, or Remote Command
Phase Specifications, Sinewave Outputs
The 5520A-PQ option has improved phase uncertainty in the normal, non-PQ, dual outputs as shown below. (See the
5520A specifications for all other output combinations.)
Output Combinations, 45 Hz to 65 Hz 1-Year Absolute Uncertainty
AC Voltage AC Voltage (Auxiliary) AC Current (LCOMP OFF)
0.65 to 3.29999 V 6.5 to 32.999 mA
6.5 to 32.9999 V 65 to 329.99 mA
65 to 329.9999 V
0.65 to 3.29999 V
0.65 to 10.9999 A
0.07 °