PQ Option
Delta (() Amplitude, Single (Sags & Swells) Function (Current)11
11-25
_______________________________________________________________________
EVTMODE <E_MODE>
(IEEE-488, RS-232, Coupled)
Set the event mode for Delta Amplitude operation
<E_MODE> is REPEAT --- Flicker operation
SINGLE --- Sag/Swell operation
Example: EVTMODE SINGLE
_______________________________________________________________________
EVTMODE?
(IEEE-488, RS-232, Sequential)
Returns the event mode in Delta Amplitude operation.
SINGLE --- for Sag/Swell operation
REPEAT --- for flicker operation
Example: EVTMODE?
_______________________________________________________________________
FLICKWAV <F_WAVE>
(IEEE-488, RS-232, Coupled)
Set the flicker modulation type; Sine or Square.
<F_WAVE> is SINE or SQUARE
Example: FLICKWAV SQUARE
Example: FLICKWAV SINE
_______________________________________________________________________
FLICKWAV?
(IEEE-488, RS-232, Sequential)
Return the flicker modulation type; Sine or Square.
_______________________________________________________________________
FUND?
(IEEE-488, RS-232, Sequential)
Returns the amplitude of the fundamental in composite harmonic mode.
Returns all zeros if not in a composite harmonic function.
1. Output amplitude
2. (Character) Units (V or A)
3. (Float) Second output amplitude for dual output functions 0 if no second amplitude)
4. (Character) Units of second amplitude (0 if no second amplitude)
Example: 1.883000E-01,A,0.000000E+00,0
_______________________________________________________________________