5522A
Operators Manual
11-24
fundamental in degrees. Numbers outside +/- 180 are automatically forced into that range
by adding or subtracting a multiple of 360.
Example: CHTONES PRI,3,33pct,0,5,20pct,0,7,16pct,0,9,11pct,0,11,9pct,0
Example: CHTONES SEC,3,0.11,-180,5,0.04,0,7,0.02,-180,0,0,0,0,0,0
_______________________________________________________________________
CHTONES? <OUTPUT CH.>
(IEEE-488, RS-232, Sequential)
Return present settings for harmonics in the format of harmonic number, amplitude, and
phase. All harmonics that are not set return to a 0 value.
<OUTPUT CH.> is PRI or SEC
Example: CHTONES? PRI
Example of return string: 2, 0.10, 30, 3, 0.02, -150,
_______________________________________________________________________
DELTAMAG <OUTPUT CH.>, <DELTA_AMPLITUDE>
(IEEE-488, RS-232, Coupled)
Set the amplitude deviation in Delta Amplitude modes (Sags/Swells & Flicker).
<OUTPUT CH.> is PRI or SEC
<DELTA_AMPLITUDE> is 0.01 % to 100.0 %
Example: DELTAMAG PRI, 0.1 deviate the Primary channel output by +10 %
Example: DELTAMAG SEC, -0.1 deviate the Primary channel output by +10 %
_______________________________________________________________________
DELTAMAG? <OUTPUT CH.>
(IEEE-488, RS-232, Sequential)
Return the amplitude deviation of the selected channel in Delta Amplitude modes.
<OUTPUT CH.> is PRI or SEC
Example: DELTAMAG? PRI
_______________________________________________________________________
DURATION <EVT_DURATION>
(IEEE-488, RS-232, Coupled)
Set the event duration in Single (Sag/Swell) Delta Amplitude mode.
<DELTA_DURATION> is .032 Sec. to 60 Sec.
Example: DURATION 49.0s --- Set the event duration to 49 Seconds
Example: DURATION 10.3 --- Set the event duration to 10.3 seconds (S is optional)
_______________________________________________________________________
DURATION?
(IEEE-488, RS-232, Sequential)
Return the event duration in Single (Sag/Swell) Delta Amplitude mode.
Example: DURATION?