SC1100 Oscilloscope Calibration Option
Verification Tables10
10-45
Table 10-16. SC1100 Option Leveled Sinewave Verification: Harmonics (cont.)
Harmonic
Nominal
Value (V p-p)
Frequency
Measured
Value (dB)
Deviation
(dB)
1-Year Spec.
(dB)
3
rd
+ harmonic 5.5 2 MHz -38
2
nd
harmonic 5.5 4 MHz -33
3
rd
+ harmonic 5.5 4 MHz -38
2
nd
harmonic 5.5 8 MHz -33
3
rd
+ harmonic 5.5 8 MHz -38
2
nd
harmonic 5.5 10 MHz -33
3
rd
+ harmonic 5.5 10 MHz -38
2
nd
harmonic 5.5 20 MHz -33
3
rd
+ harmonic 5.5 20 MHz -38
2
nd
harmonic 5.5 40 MHz -33
3
rd
+ harmonic 5.5 40 MHz -38
2
nd
harmonic 5.5 80 MHz -33
3
rd
+ harmonic 5.5 80 MHz -38
2
nd
harmonic 5.5 100 MHz -33
3
rd
+ harmonic 5.5 100 MHz -38
2
nd
harmonic 5.5 200 MHz -33
3
rd
+ harmonic 5.5 200 MHz -38
2
nd
harmonic 5.5 400 MHz -33
3
rd
+ harmonic 5.5 400 MHz -38
2
nd
harmonic 5.5 600 MHz -33
3
rd
+ harmonic 5.5 600 MHz -38
2
nd
harmonic 3.5 1000 MHz -33
3
rd
+ harmonic 3.5 1000 MHz -38