Front Panel Operation
Sample Applications 4
4-59
AC Current Frequency
35.0 mA 1.0 kHz
350.0 mA 60 Hz
350.0 mA 1.0 kHz
i. Press Y on the calibrator and switch the DMM function switch to w.
j. Set the calibrator output to 350 μA at 0 Hz. and press O. Verify the error is
within specification.
k. Repeat the previous step using 3500 μA at 0 Hz.
l. Press Y on the calibrator and press the blue key on the DMM to switch to ac
measurements.
m. Set the calibrator output to 350.0 μA at 60 Hz and press O. Verify the error is
within specification.
n. Repeat the previous step with the following calibrator settings:
AC Current Frequency
350.0 μA 1.0 kHz
3500.0 μA 60 Hz
3500.0 μA 1.0 kHz
11. Test the High current function.
a. Press R on the calibrator.
b. Verify that the calibrator is in standby and connect the DMM as shown in
Figure 4-20.
5522A
CALIBRATOR
87
TRUE RMS MULTIMETER
MIN MAX
RANGE HOLD
H
REL Hz
mA
A
A
m
mV
V
V
OFF
A
COM
V
mA
m
A
PEAK MIN MAX
gjh027.eps
Figure 4-20. Cable Connections for Testing an 80 Series High Amps Function
c. Set the calibrator output to 3.5 A at 0 Hz and press O. Verify the error is
within specification.
d. Repeat the previous step using 10.0 A at 0 Hz. Verify the error is within
specification.
e. Press Y on the calibrator and press the blue key on the DMM to switch to ac
measurements.