Operators Manual
6-12
CUR_POST
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(Current Post command) Selects the binding posts for current output. This also applies to
power outputs. The current post setting is retained until the power is turned off or the
R button is pressed.
Parameters: AUX (selects the AUX terminals)
A20 (selects the 20A terminals)
Example: CUR_POST AUX
Selects the Calibrator front panel AUX terminals for the output current.
CUR_POST?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(Current Post query) Returns the active front panel binding post terminals used for
current output: AUX or 20A.
Responses: AUX (AUX terminals are selected)
A20 (20A terminals are selected)
Example: CUR_POST? returns AUX
Returns AUX when the AUX terminals are selected for output current.
DAMPEN
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Dampen Mode for Pressure Measurement command) Activates or deactivates
dampening (averaging) of pressure readings.
Parameter: ON (dampen on)
OFF (dampen off)
Example: DAMPEN ON
DAMPEN?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Dampen Mode for Pressure Measurement query) Returns whether dampening
(averaging) of pressure readings is active.
Response: (character) ON (dampen on)
(character) OFF (dampen off)
Example: DAMPEN returns ON
DBMZ
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(dBm Impedance command) Sets the impedance used for dBm outputs (ac volts).
Parameters: Z50 (50 ohms)
Z75 (75 ohms)
Z90 (90 ohms)
Z100 (100 ohms)
Z135 (135 ohms)
Z150 (150 ohms)
Z300 (300 ohms)
Z600 (600 ohms)