Remote Commands
Commands
6
6-19
Features
Temperature
Standard
its-90 Display Contrast* level 7,7
Host Connection gpib (IEEE-488) Display Brightness* level 1,0
GPIB Port Address 4 RTD Power Up
Default Type
pt385
Serial Ports 8 bits, 1 stop bit, xon/xoff,
parity none, 9600 baud
Thermocouple
Power Up Default
Type
K
EOL (end of line) CRLF Current Limits ±20.5 A
EOF (end of file) 012,000 Voltage Limits ±1020 V
Remote I/F term
Remote commands (see Chapter 6)
SRQSTR SRQ: %02x %02x %04x
%04x
*PUD string cleared
* Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.
Defaults
Reference Clock Internal Reference Phase 0°
dBm Impedance 600 Ω Pressure Unit PSI
FUNC?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Function query) Returns the present output, measurement, or calibration function. See
the response below for output & measurement modes.
Responses: DCV (dc volts function)
ACV (ac volts function)
DCI (dc current function)
ACI (ac current function)
RES (ohms function)
CAP (capacitance function)
RTD (temperature with an rtd function)
TC_OUT (temperature with a thermocouple function)
DC_POWER (dc power function)
AC_POWER (ac power function)
DCV_DCV (dual dc volts function)
ACV_ACV (dual ac volts function)
TC_MEAS (measure temperature with a thermocouple)
SACV (oscilloscope ac volts function)
SDCV (oscilloscope dc volts function)
MARKER (oscilloscope marker function)
LEVSINE (oscilloscope leveled sine function)
EDGE (oscilloscope edge function)
Example: FUNC? returns DCV_DCV
Returns DCV_DCV when the Calibrator output function dual dc volts.