5522A
Operators Manual
6-22
Example: ISCE1 6272
Load decimal 6272 (binary 0001010001000000) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT).
ISCE1?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status 0 to 1 Change Enable query) Returns the two bytes from the 16-bit
ISCE1 register. (See “Instrument Status Change Enable Registers” in Chapter 5 for more
information.)
Response: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISCE1? returns 6272
Returns decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCR?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status Change Register query) Returns and clears the contents of the
Instrument Status 1 to 0 Change Register (ISCR0) and Instrument Status 0 to 1 Change
Register (ISCR1). (See “Instrument Status Change Register” in Chapter 5 for more
information.)
Response: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISCR? returns 6272
Returns decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCR0?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status 1 to 0 Change Register query) Returns and clears the contents of the
Instrument Status 1 to 0 Change Register.
Response: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISCRO? returns 6272
Returns decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCR1?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status 0 to 1 Change Register query) Returns and clears the contents of the
Instrument Status 0 to 1 Change Register.
Response: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISCR1? returns 6272
Returns decimal 6272 (binary 0001010001000000) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISR?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Instrument Status Register query) Returns contents of the Instrument Status Register.
Response: <value> (decimal equivalent of the 16 bits, 0 to 32767)
Example: ISR? returns 6272

Table of Contents

Related product manuals