5522A
Operators Manual
6-26
*OPC?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Operations Complete query) Returns a 1 after all pending operations are complete. This
command causes program execution to pause until operations are complete. (See *WAI.)
Response: 1 (all operations are complete)
Example: *OPC? returns 1
Returns 1 when all pending operations are complete.
OPER
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Operate command) Activates the Calibrator output if it is in standby. This is the same as
pressing the Calibrator front panel O key. If there are errors in the error queue, the
OPER command is inhibited for outputs 33 V and over. (Also see the ERR? command
and STBY command.)
Parameter: (None)
Example: OPER
Connect the selected output to the Calibrator front panel terminals. Also lights the
annunciator in the O key.
OPER?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Operate query) Returns the operate/standby setting.
Response: 1 (Operate)
0 (Standby)
Example: OPER? returns 1
Returns 1 when the Calibrator is in operate.
*OPT?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Options command) Returns a list of the installed hardware and software options.
Responses: <option string>,<option string>,... (options list, separated by commas)
0 (no options are installed)
Example: *OPT? returns SC600
Returns SC600 when the Oscilloscope Calibration Option is installed.
OUT
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(Output command) Sets the output of the Calibrator and establishes a new reference
point for the error mode. If only one amplitude is supplied, the Calibrator sources a single
output. If two amplitudes are supplied, the Calibrator sources two outputs. The second
amplitude will be sourced at the AUX terminals for dual voltage outputs. If the frequency
is not supplied, the Calibrator will use the frequency that is presently in use.
To source or measure a temperature, select the desired sensor and sensor parameters first.
(See the TSENS_TYPE, RTD_*, and TC_* commands.)
To source a signal using the Calibrator scope options, refer to the SCOPE command in
Chapter 8.