Remote Commands
Commands
6
6-33
Example: RANGELCK? returns OFF
Returns OFF when the range for dc volts or dc current is not locked (autoranging
enabled).
REFCLOCK
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Reference Clock command) Sets the reference clock source (internal or through the 10
MHz IN BNC connector).
Parameter INT (Sets internal reference clock)
EXT (Sets external reference clock)
Example: REFCLOCK INT
Once set, the Calibrator retains the external guard setting until power off or reset.
REFCLOCK?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Reference Clock query) Returns the reference clock source (internal or through the 10
MHz IN BNC connector).
Response: (character) INT (Reference clock is internal)
(character) EXT (Reference clock is external)
Example: REFCLOCK? returns INT
REFCLOCK_D
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Reference Clock Default command) Sets the power-up and reset default for the
reference clock source (internal or through the 10 MHz IN BNC connector).
Parameters INT (Sets internal reference clock)
EXT (Sets external reference clock)
The reference clock is set to the default at power on, reset, and when going into an ac
function.
Example: REFCLOCK_D INT
REFCLOCK_D?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Reference Clock Default query) Returns the power-up and reset default for the
reference clock source (internal or through the 10 MHz IN BNC connector).
Responses: (character) INT (Reference clock is internal)
(character) EXT (Reference clock is external)
Example: REFCLOCK_D? returns INT
REFOUT?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Reference Output query) Returns the present value of the reference when editing the
output (error mode). If not editing the output using the INCR command, the return is 0
(0E+00). The reference value is set with the OUT, NEWREF or MULT commands. To
determine which quantity is being edited, use the EDIT? and OUT? commands.
Response: <reference value>
Example: REFOUT? returns 0E+00