5522A
Operators Manual
7-16
Table 7-10. Verification Tests for Capacitance (cont.)
Range Output
Test Frequency
or Current
Lower Limit Upper Limit
3.2999 nF 2.0000 nF 1 kHz 1.9824 nF 2.0176 nF
10.9999 nF 7.0000 nF 1 kHz 6.9767 nF 7.0233 nF
10.9999 nF 10.9000 nF 1 kHz 10.8693 nF 10.9307 nF
32.9999 nF 20.0000 nF 1 kHz 19.8620 nF 20.1380 nF
109.999 nF 70.000 nF 1 kHz 69.767 nF 70.233 nF
109.999 nF 109.000 nF 1 kHz 108.693 nF 109.307 nF
329.999 nF 200.000 nF 1 kHz 199.320 nF 200.680 nF
329.999 nF 300.000 nF 1 kHz 299.130 nF 300.870 nF
1.09999 μF 0.70000 μF 100 Hz 0.69767 μF 0.70233 μF
1.09999 μF 1.09000 μF 100 Hz 1.08693 μF 1.09307 μF
3.29999 μF 2.00000 μF 100 Hz 1.99320 μF 2.00680 μF
3.29999 μF 3.00000 μF 100 Hz 2.99130 μF 3.00870 μF
10.9999 μF 7.0000 μF 100 Hz 6.9767 μF 7.0233 μF
10.9999 μF 10.9000 μF 100 Hz 10.8693 μF 10.9307 μF
32.9999 μF 20.0000 μF 100 Hz 19.9100 μF 20.0900 μF
32.9999 μF 30.0000 μF 100 Hz 29.8800 μF 30.1200 μF
109.999 μF 70.000 μF 50 Hz 69.662 μF 70.338 μF
109.999 μF 109.000 μF 50 Hz 108.529 μF 109.471 μF
329.999 μF 200.000 μF 54 μA dc 199.020 μF 200.980 μF
329.999 μF 300.000 μF 80 μA dc 298.680 μF 301.320 μF
1.09999 mF 0.33000 mF 90 μA dc 0.32788 mF 0.33212 mF
1.09999 mF 0.70000 mF 180 μA dc 0.69662 mF 0.70338 mF
1.09999 mF 1.09000 mF 270 μA dc 1.08529 mF 1.09471 mF
3.2999 mF 1.1000 mF 270 μA dc 1.0933 mF 1.1067 mF
3.2999 mF 2.0000 mF 540 μA dc 1.9902 mF 2.0098 mF
3.2999 mF 3.0000 mF 800 μA dc 2.9868 mF 3.0132 mF
10.9999 mF 3.3000 mF 900 μA dc 3.2788 mF 3.3212 mF
10.9999 mF 10.9000 mF 2.7 mA dc 10.8529 mF 10.9471 mF
32.9999 mF 20.0000 mF 5.4 mA dc 19.8300 mF 20.1700 mF
32.9999 mF 30.0000 mF 8.0 mA dc 29.7600 mF 30.2400 mF
110.000 mF 33.000 mF 9.0 mA dc 32.570 mF 33.430 mF
110.000 mF 110.000 mF 27.0 mA dc 108.800 mF 111.200 mF