SC600 Oscilloscope Calibration Option
Remote Commands and Queries 9
9-29
SCOPE?
(IEEE-488, RS-232, Sequential)
Returns the oscilloscope’s current mode of operation. Returns OFF if the oscilloscope is
off.
Parameters: None
Response: <character> (Returns OFF, VOLT, EDGE, LEVSINE, MARKER, or
WAVEGEN.)
TRIG
(IEEE-488, RS-232, Sequential)
Programs the oscilloscope’s trigger output BNC.
Parameters: OFF (Turns the trigger output off.)
DIV1 (Turns the trigger output on. Frequency is the same as the
signal at SCOPE output.)
DIV10 (Turns the trigger output on. Frequency is 1/10 of the signal at
SCOPE output.)
DIV100 (Turns the trigger output on. Frequency is 1/100 of the signal
at SCOPE output.)
TRIG?
(IEEE-488, RS-232, Sequential)
Returns the output setting of the oscilloscope’s trigger.
Parameters: (None)
Response: <character> (Returns OFF, DIV1, DIV10, or DIV100.)
OUT_IMP
(IEEE-488, RS-232, Sequential)
Programs the oscilloscope’s output impedance.
Parameters: Z50 (Programs the oscilloscope’s output impedance to 50 Ω.)
Z1M (Programs the oscilloscope’s output impedance to 1 MΩ.)
OUT_IMP?
(IEEE-488, RS-232, Sequential)
Returns the impedance setting of the oscilloscope’s output.
Parameters: (None)
TRIG
Programs the instrument range in PULSE, MEAS Z, and OVERLD modes.
Parameters: TP0DB Sets the range to 2.5 V in pulse mode.
TP8DB Sets the range to 1.0 V in pulse mode.
TP20DB Sets the range to 250 mV in pulse mode.
TP28DB Sets the range to 100 mV in pulse mode.
TZ50OHM Sets the impedance to 50 in Meas Z mode.