5522A
Operators Manual
9-44
Table 9-10. SC600 Option Leveled Sine Wave Verification: Flatness (cont.)
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.07 600 MHz 0.0029
0.099 50 kHz na na
0.099 30 MHz 0.001585
0.099 70 MHz 0.001585
0.099 120 MHz 0.00208
0.099 290 MHz 0.00208
0.099 360 MHz 0.00406
0.099 390 MHz 0.00406
0.099 400 MHz 0.00406
0.099 480 MHz 0.00406
0.099 570 MHz 0.00406
0.099 580 MHz 0.00406
0.099 590 MHz 0.00406
0.099 600 MHz 0.00406
0.1 50 kHz na na
0.1 30 MHz 0.0016
0.1 70 MHz 0.0016
0.1 120 MHz 0.0021
0.1 290 MHz 0.0021
0.1 360 MHz 0.0041
0.1 390 MHz 0.0041
0.1 400 MHz 0.0041
0.1 480 MHz 0.0041
0.1 570 MHz 0.0041
0.1 580 MHz 0.0041
0.1 590 MHz 0.0041
0.1 600 MHz 0.0041
0.25 50 kHz na na
0.25 30 MHz 0.00385
0.25 70 MHz 0.00385
0.25 120 MHz 0.0051
0.25 290 MHz 0.0051

Table of Contents