Introduction and Specifications
Detailed Specifications 1
1-15
AC Current (Sine Wave) (cont.)
Absolute Uncertainty,
tcal ±5 °C
±(% of output + μA)
Range Frequency
90 days 1 year
Max Distortion &
Noise 10 Hz to
100 kHz BW
±(% of output + floor)
Max
Inductive
Load μH
LCOMP On
10 to 100 Hz 0.2 + 0.2 0.25 + 0.2 0.1 + 1.0
29.00 to
329.99 μA
100 Hz to 1 kHz 0.5 + 0.5 0.6 + 0.5 0.05 + 1.0
10 to 100 Hz 0.2 + 0.3 0.25 + 0.3 0.15 + 1.5
0.33 to
3.29999 mA
100 Hz to 1 kHz 0.5 + 0.8 0.6 + 0.8 0.06 + 1.5
10 to 100 Hz 0.07 + 4 0.08 + 4 0.15 + 5
3.3 to
32.9999 mA
100 Hz to 1 kHz 0.18 + 10 0.2 + 10 0.05 + 5
10 to 100 Hz 0.07 + 40 0.08 + 40 0.15 + 50
33 to
329.999 mA
100 Hz to 1 kHz 0.18 + 100 0.2 + 100 0.05 + 50
10 to 100 Hz 0.1 + 200 0.12 + 200 0.2 + 500
0.33 to
2.99999 A
100 to 440 Hz 0.25 + 1000 0.3 + 1000 0.25 + 500
400
10 to 100 Hz
0.1 + 2000
[2]
0.12 + 2000
[2]
0.1 + 0
3 to 20.5 A
[1]
100 Hz to 1 kHz
0.8 + 5000
[3]
1.0 + 5000
[3]
0.5 + 0
400
[4]
[1] Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure B. The current may be provided 60-
T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output
current in Amps. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5522A is
outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those
conditions, the allowable "on" time indicated by the formula and Figure B is achieved only after the 5522A is outputting currents
<5 A for the "off" period first.
[2] For currents >11 A, Floor specification is 4000 μA within 30 seconds of selecting operate. For operating times >30 seconds,
the floor specification is 2000 μA.
[3] For currents >11 A, Floor specification is 1000 μA within 30 seconds of selecting operate. For operating times >30 seconds,
the floor specification is 5000 μA.
[4] Subject to compliance voltages limits.
Range
Resolution μA
Max Compliance Voltage V rms [1]
0.029 to 0.32999 mA 0.01 7
0.33 to 3.29999 mA 0.01 7
3.3 to 32.9999 mA 0.1 5
33 to 329.999 mA 1 5
0.33 to 2.99999 A 10 4
3 to 20.5 A 100 3
[1] Subject to specification adder for compliance voltages greater than 1 V rms.