SC1100 Oscilloscope Calibration Option
Verification Tables10
10-47
Table 10-17. SC1100 Option Leveled Sinewave Verification: Flatness (cont.)
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
0.0099 360 MHz 0.000496
0.0099 390 MHz 0.000496
0.0099 400 MHz 0.000496
0.0099 480 MHz 0.000496
0.0099 570 MHz 0.000496
0.0099 580 MHz 0.000496
0.0099 590 MHz 0.000496
0.0099 600 MHz 0.000496
0.01 10 MHz na na
0.01 30 MHz 0.00025
0.01 70 MHz 0.00025
0.01 120 MHz 0.0003
0.01 290 MHz 0.0003
0.01 360 MHz 0.0005
0.01 390 MHz 0.0005
0.01 400 MHz 0.0005
0.01 480 MHz 0.0005
0.01 570 MHz 0.0005
0.01 580 MHz 0.0005
0.01 590 MHz 0.0005
0.01 600 MHz 0.0005
0.01 1000 MHz 0.0005
0.025 10 MHz na na
0.025 30 MHz 0.000475
0.025 70 MHz 0.000475
0.025 120 MHz 0.0006
0.025 290 MHz 0.0006
0.025 360 MHz 0.0011
0.025 390 MHz 0.0011
0.025 400 MHz 0.0011
0.025 480 MHz 0.0011
0.025 570 MHz 0.0011