5522A
Operators Manual
10-52
Table 10-17. SC1100 Option Leveled Sinewave Verification: Flatness (cont.)
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
0.8 600 MHz 0.0321
0.8 1000 MHz 0.0321
1.2 10 MHz na na
1.2 30 MHz 0.0181
1.2 70 MHz 0.0181
1.2 120 MHz 0.0241
1.2 290 MHz 0.0241
1.2 360 MHz 0.0481
1.2 390 MHz 0.0481
1.2 400 MHz 0.0481
1.2 480 MHz 0.0481
1.2 570 MHz 0.0481
1.2 580 MHz 0.0481
1.2 590 MHz 0.0481
1.2 600 MHz 0.0481
1.2 1000 MHz 0.0481
1.3 10 MHz na na
1.3 30 MHz 0.0196
1.3 70 MHz 0.0196
1.3 120 MHz 0.0261
1.3 290 MHz 0.0261
1.3 360 MHz 0.0521
1.3 390 MHz 0.0521
1.3 400 MHz 0.0521
1.3 480 MHz 0.0521
1.3 570 MHz 0.0521
1.3 580 MHz 0.0521
1.3 590 MHz 0.0521
1.3 600 MHz 0.0521
1.3 1000 MHz 0.0521
3.4 10 MHz na na
3.4 30 MHz 0.0511

Table of Contents