5522A
Operators Manual
3-22
Table 3-3. Factory Defaults for SETUP Menus Power-Up Defaults
Parameter Setting
SETUP Menu
(Figure 3-4.)
User report string (*PUD
string)
Cleared. D
Error units > 0.1% G1
SC-600 option overload test
safety timeout
10 s G1
Temperature standard its-90 G1
Host interface gpib (IEEE-488) G1
UUT serial interface 8 bits, 1 stop bit, xon/xoff, parity none,
9600 baud
I
Host serial interface term, 8 bits, 1 stop bit, xon/xoff, parity
none, 9600 baud, CRLF, 012,000
K, L, M, N
GPIB Port Address 4 O
Display brightness (Note) level 1,0 P
Display contrast (Note) level 7,7 P
dBm impedance 600 Ω S
Pressure units psi S
RTD type pt385 S1
Thermocouple type K S1
Phase reference 0.00° S3
10 MHz reference clock internal S2
Current limits ±20.5 A U
Voltage limits ±1020 V V
Note:
Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6, and 7