Operators Manual
4-20
nn227f.eps
Note
At voltage outputs of 100 V and above (nominal), you may notice a slight
high-pitched sound. This is normal.
1. Press a multiplier key, if necessary. For example, press c.
2. Output in volts. Press V.
Output in dBm. Press b V. Select an impedance for dBm from a list on the
Control Display using the rightmost softkey.
3. The Control Display now shows the amplitude of your entry. For example,
2.44949 V (below).
nn073f.eps
4. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. The press the H key. For example, 1.1234 kHz
(below).
nn074f.eps
5. Press E . The calibrator clears your entry from the Control Display and copies it
into the Output Display (below is typical).