Page 1 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Volume Apnea On On On On
.
Page 2 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Vent Mode VCV PCV VCV VCV
TV 500 150 500 500
RR 12 16 12 12
RR 12 16 12 12
Minimum RR 12 16 12 12
I:E 1:2 1:2 1:2 1:2
Tpause Off Off Off Off
PEEP Off Off Off Off
.
Page 3 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Pmax 40 40 40 40
Pinsp 5 5 5 5
Tinsp 1.7 1.7 1.7 1.7
Trig Window 25 25 25 25
Flow Trigger 2 2 2 2
Psupport 2 20 2 2
End of Breath 30 30 30 30
.
Page 4 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Backup Time 30 30 30 30
Rise Rate Auto Auto Auto Auto
Rise Rate 5 5 5 5
PSV Rise Rate Auto Auto Auto Auto
Exit Backup 2 2 2 2
.
Page 5 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Sweep Speed Fast Fast Fast Fast
Top Waveform Paw Paw Paw Paw
Middle Waveform Flow Flow Flow Flow