Page 5 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
Bottom Waveform CO2 CO2 CO2 CO2
Split Screen Paw Paw Paw Paw
Left Digit Field Loops Loops Loops Loops
Middle Left Digit
Field
Resp Resp Resp Resp
Middle Right Digit
Field
Agent Agent Agent Agent
Right Digit Field Gases Gases Gases Gases
.
Page 6 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
MV Low 2 2 Off 2
MV High 10 10 Off 10
TV Low Off Off Off Off
TV High 1000 1000 Off 1000
RR Low Off Off Off Off
RR High Off Off Off Off
EtCO2 Low 3 3 3 3
EtCO2 High 8 8 8 8
EtO2 Low Off Off Off Off
EtO2 High Off Off Off Off
FiCO2 High Off Off Off Off
FiO2 Low 21 21 21 21
FiO2 High Off Off Off Off
.
Page 7 Default settings for default case types
Settings ADULT PEDIATRIC LOCAL CUSTOM 1
EtIso Low Off Off Off Off
EtIso High Off Off Off Off
FiIso Low Off Off Off Off
FiIso High 5 5 5 5
EtSev Low Off Off Off Off
EtSev High Off Off Off Off
FiSev Low Off Off Off Off
FiSev High 8 8 8 8
EtDes Low Off Off Off Off
EtDes High Off Off Off Off
FiDes Low Off Off Off Off
FiDes High 15 15 15 15
12 Super user mode
2067226-001 12-13