67
7
Maintenance
General
The modular structure of the FVP110 increases the ease of maintenance work. This section
describes cleaning and part replacement procedures.
The FVP110 is a precision instrument; read the following carefully when carrying out
maintenance. For calibrations, see “Setup” on page 59
Precautions for CENELEC, and JIS flameproof type instruments:
Flameproof type instruments must be, as a rule, removed to a
non-hazardous area for maintenance and be disassembled and
reassembled to the original state. For details, see “Installation and
Operating Precautions for JIS Flameproof Equipment” on page 265.
On flameproof type instruments the terminal cover is locked by an
Allen head bolt (shrouding bolt). When a shrouding bolt (Figure 39) is
driven clockwise by an Allen wrench, it goes in and cover lock is
released, and the cover can be opened. When a cover is closed it
must be locked by a shrouding bolt. Tighten the shrouding bolt to a
torque of 0.7 N·m.
Figure 39 Shrouding Bolts