GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–42
REV 6
4–4–2 Power Supply Adjustment Procedure (continued)
TABLE 4–3
POWER SUPPLY 2 MEASUREMENT LOCATION
POWER SUPPLY MEASURE AT RETURN AT
+5V for Digital
+12V
+5V for Analog
–5V
+5V Terminal on Power Supply
See ILLUSTRATION 4–32
+5V Terminal on Power Supply
See ILLUSTRATION 4–32
+12V Terminal on Power Supply
See ILLUSTRATION 4–32
–5V Terminal on Power Supply
See ILLUSTRATION 4–32
GND Terminal on Power Supply
See ILLUSTRATION 4–32
GND Terminal on Power Supply
See ILLUSTRATION 4–32
GND Terminal on Power Supply
See ILLUSTRATION 4–32
GND Terminal on Power Supply
See ILLUSTRATION 4–32
+15V
+15V Terminal on Power Supply
See ILLUSTRATION 4–32
GND Terminal on Power Supply
See ILLUSTRATION 4–32
TABLE 4–4
POWER SUPPLY 2 MEASUREMENT TOLERANCES
+12.10 V
–5.10 V
+5.35 V
POWER SUPPLY MIN. MAX
+5.25 V+5V for Digital
+12V
+5V for Analog
–5V
+12.20 V
–5.20 V
+5.45 V
+5.35 V
+15.40 V
+15V
+15.50 V