GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL
2127661
RENEWAL PARTS
6–128
REV 9
6–2–32 Side Cover Right (FRU No. 302)
Time Required
5 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–54.
1. Turn OFF the system.
2. Remove the Front Base Cover (FRU No. 305). Refer to 6–2–35 on page 6–131.
3. Remove the two screw caps (located at
3
and
4
) and unscrew five screws (
1
5
5
).
4. Remove the Rear Door Assy. Refer to 6–2–34 on page 6–130.
5. Remove the Side Cover Right.
SIDE COVER RIGHT DISASSEMBLY
ILLUSTRATION 6–54