GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL
2127661
RENEWAL PARTS
6–86
REV 9
6–2–10 CRT Assy (FRU No. 115)
Time Required
0.5 Hours
Tool Required
Screwdriver
Hexagonal Wrench
Procedure
WARNING!
PERSONAL INJURY HAZARD. THE NECK ASSY IS SPRING LOADED. RELEASING ARM
WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVEMENT. KEEP
YOUR HEAD AND BODY AWAY FROM ABOVE NECK ASSY.
Refer to ILLUSTRATION 6–26 on page 6–87.
1. Turn OFF the system.
2. Remove the Monitor Assy (FRU 100). Refer to 6–2–1 on page 6–70.
3. Remove the Monitor Cover Set (
1
: FRU 101). Refer to 6–2–2 on page 6–72.
4. Remove the Speaker Assy (
2
: FRU 112). Refer to 6–2–8 on page 6–82.
5. Remove the tilt mechanism.