GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL
2127661
RENEWAL PARTS
6–79
REV 9
6–2–6 CRT Filter (FRU No. 109), CRT Filter Clamp Set (FRU No. 110)
Time Required
1 Minutes
Tool Required
Not necessary
Procedure
Refer to ILLUSTRATION 6–22.
1. Turn OFF the system.
2. Slide the left side filter clamp leftwards and pull it out.
3. Slide the right side filter clamp rightwards and pull it out.
4. Remove the CRT Filter.
CRT FILTER REMOVAL
ILLUSTRATION 6–22
Note
The CRT Filter (FRU109) includes the CRT Filter Clamps (FRU110).