EasyManua.ls Logo

GE LOGIQ 400 - Escutcheon Assy (FRU No. 153)

GE LOGIQ 400
526 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS LOGIQ 400 SERVICE MANUAL
2127661
RENEWAL PARTS
6–93
REV 9
6–2–14 Escutcheon Assy (FRU No. 153)
Time Required
10 Minute
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–30.
1. Turn OFF the system.
2. Remove the Monitor Cover (FRU 152). Refer to 6–2–13 on page 6–92.
3. Unscrew four screws (
1
4
).
4. Remove the Escutcheon Assy (FRU 153).
Note
When replacing the Escutcheon with the new Escutcheon Assy (2164507), Be sure to attach the
proper label included with the Escutcheon Assy on to the Escutcheon. To attach the new label, refer to
the procedures described in the following pages.
ESCUTCHEON ASSY REMOVAL
ILLUSTRATION 6–30

Table of Contents

Other manuals for GE LOGIQ 400

Related product manuals