LOGIQ 400 SERVICE MANUALGE MEDICAL SYSTEMS
2127661
RENEWAL PARTS
6–138
REV 9
6–2–41 Probe Holder (FRU No. 312)
Time Required
1 Minutes
Tool Required
Not necessary
Procedure
Refer to ILLUSTRATION 6–64.
1. Turn OFF the system.
2. Pull the latch located at bottom of the probe holder.
3. Pull the Probe Holder out.
PROBE HOLDER REMOVAL
ILLUSTRATION 6–64