6–4–3 Initializing Hard Disk
2. Type the following command and press the return key.
VTSH>mediainit_0.0_PHILE_240000000_20_10000_4000
Note
For the console with software version 3.40 or later, use the following command:
VTSH>mediainit_0.0_PHILE_540000000_20_10000_4000
Note
The character “_” described in operating commands indicates the space.
After 20 minutes, the system completes the formatting of hard disk. The message ”***Format Completed***” is
displayed on the screen of Laptop PC.
3. Type the following command and press the return key. The hard disk become valid by entering this command.
VTSH>mount_0.0
6–4–4 Installing Software
1. Insert the appropriate system software MO disk into the MO disk drive of LOGIQ 400.
2. Type the following command and press the return key. The system will start writing software in the hard disk.
VTSH>cp_–r_2.0/*_0.0/
3. Turn the main switch of the system OFF after completing to write the software into the hard disk.
4. Turn the Laptop PC power OFF.
5. Disconnect the interface cable from the system.