LOGIQ 400 SERVICE MANUAL
TM
8–57 OPTIONS
REV 5
8–5–8 Operational Check–out
1. Connect the ECG patient cable with the connector on the ECG panel as shown in ILLUSTRATION 8–62.
CONNECTING ECG PATIENT CABLE
ILLUSTRATION 8–62
2. Connect the electrodes of ECG patient cable with the ECG simulator.
3. Plug in the system and turn the system power ON.