GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
RENEWAL PARTS
6–171
REV 9
6–2–65 ECG Board Assy (FRU No. 506)
Time Required
15 Minutes
Tool Required
Screwdriver
Procedure
Refer to ILLUSTRATION 6–89.
1. Turn OFF the system.
2. Remove the Side Cover Right (FRU 302). Refer to 6–2–32 on page 6–128.
3. Remove the ECG Panel Assy (FRU 511). Refer to 6–2–66 on page 6–172.
4. Disconnect the connector (
1
).
5. Unscrew two screws (
2
and
3
).
6. Remove the ECG Board Assy.
ECG BOARD ASSY DISASSEMBLY
ILLUSTRATION 6–89
Note
New type of ECG Board Assy (2207806) has the reserved connector (P69). This connector is not
used for the LOGIQ 400 console.